Focus on SURFACAL - Calibrate your surface sensors!

The main advantage of the SURFACAL method over previous methods is that you can calibrate surface sensors in conditions identical to their operating conditions.

SURFACAL is a BNE-LNE surface temperature reference for calibrating surface sensors in the laboratory or on site.
The main advantage of the SURFACAL method over previous methods is that you can calibrate surface sensors in conditions identical to their operating conditions.


# Calibration by comparison with the national references (LNE in France)
# Calibration of Pt100 surface sensors (2,3 or 4 wires) and thermocouples (as long as their specifications are metrologically compatible)
# Temperature displayed on front panel: +35°C to +300°C. Resolution: 0.1°C
# Accuracy: +/-0.6°C to +/-2.8°C
# Supplied with LNE Calibration Certificate at 2 points, 50°C and 300°C
# 316L stainless-steel surface as standard. Diameter: 90mm
# Aluminium surface available as an option
# Operation in horizontal or vertical position
# Transportable for use on site

Ask for further information